# LiteVNA calibration
The calibration procedure is as follows:
1. Reset current calibration state. Select menu item CAL →RESET and then
→CALIBRATE.
2. Connect OPEN standard to port 1 cable and click →OPEN. Wait for the current
calibration step to be completed, and the cursor will automatically move to the
next step.
3. Connect SHORT standard to port 1 cable and click →SHORT. Wait for the
current calibration step to be completed, and the cursor will automatically move
to the next step.
4. Connect LOAD standard to port 1 cable and click →LOAD. Wait for the current
calibration step to be completed, and the cursor will automatically move to the
next step.
11
5. Connect LOAD standard to port 1 cable and click →ISOLN. Wait for the current
calibration step to be completed, and the cursor will automatically move to the
next step.
6. Connect the THRU standard between the port 1 and port 2 cable ends, and
click
→THRU. Wait for the current calibration step to be completed, and the cursor
will automatically move to the next step.
7. Click →DONE.
8. In the pop-up save menu inside choose the location you need to save, save 0
will be automatically called back in the next boot.# LiteVNA calibration
The calibration procedure is as follows:
1. Reset current calibration state. Select menu item CAL →RESET and then
→CALIBRATE.
2. Connect OPEN standard to port 1 cable and click →OPEN. Wait for the current
calibration step to be completed, and the cursor will automatically move to the
next step.
3. Connect SHORT standard to port 1 cable and click →SHORT. Wait for the
current calibration step to be completed, and the cursor will automatically move
to the next step.
4. Connect LOAD standard to port 1 cable and click →LOAD. Wait for the current
calibration step to be completed, and the cursor will automatically move to the
next step.
11
5. Connect LOAD standard to port 1 cable and click →ISOLN. Wait for the current
calibration step to be completed, and the cursor will automatically move to the
next step.
6. Connect the THRU standard between the port 1 and port 2 cable ends, and
click
→THRU. Wait for the current calibration step to be completed, and the cursor
will automatically move to the next step.
7. Click →DONE.
8. In the pop-up save menu inside choose the location you need to save, save 0
will be automatically called back in the next boot.# LiteVNA calibration
The calibration procedure is as follows:
1. Reset current calibration state. Select menu item CAL →RESET and then
→CALIBRATE.
2. Connect OPEN standard to port 1 cable and click →OPEN. Wait for the current
calibration step to be completed, and the cursor will automatically move to the
next step.
3. Connect SHORT standard to port 1 cable and click →SHORT. Wait for the
current calibration step to be completed, and the cursor will automatically move
to the next step.
4. Connect LOAD standard to port 1 cable and click →LOAD. Wait for the current
calibration step to be completed, and the cursor will automatically move to the
next step.
11
5. Connect LOAD standard to port 1 cable and click →ISOLN. Wait for the current
calibration step to be completed, and the cursor will automatically move to the
next step.
6. Connect the THRU standard between the port 1 and port 2 cable ends, and
click
→THRU. Wait for the current calibration step to be completed, and the cursor
will automatically move to the next step.
7. Click →DONE.
8. In the pop-up save menu inside choose the location you need to save, save 0
will be automatically called back in the next boot.