This is a quick step-by-step protocol for using the Qubit fluorometer with either the BR (broad range) or HS (high sensitivity) dsDNA kits. Qubit provides a much more accurate reading of DNA concentration than the Nanodrop spectrophotometer and is an essential tool for library preparation.
Reagent/Assay | Assay range | Sample starting concentration range |
---|---|---|
Qubit dsDNA HS Assay | 0.2–100 ng | 10 pg/μl–100 ng/μl |
Qubit dsDNA BR Assay | 2–1,000 ng | 100 pg/μl–1 μg/μl |
Protocol
Pull out Qubit standards from fridge and allow to warm to RT (~5 min). Mix by gently inverting each container several times.
Pull out Qubit buffer and reagent (both stored at RT). Gently mix buffer by inverting several times. Reagent should be stored in dark box and exposed to light as little as possible.
Combine buffer and reagent into an appropriate-sized tube and vortex for 10 sec. This is your Qubit mix.
Pull out enough Qubit tubes for your samples plus 2 standards. Leave the lids open. Make sure to use the correct Qubit tubes. These need to be optically clear, don't autoclave them!
Pipet 190 μl of Qubit mix into the 2 standard tubes. It is not necessary to change tips between tubes.
Pipet 10 μL of standard 1 into one of the standard tubes and pipet 10μL of standard 2 into the other. Close and label the standard tubes “S1” and “S2” or any other code. Suggestion: use a simple dot or double dot for those.
Pipet 198 μl of Qubit mix into each of the sample tubes. It is not necessary to change tips between tubes.
Using a separate filter tip, pipet 2 μL of sample into the sample tubes. Make sure to insert the pipet tip into the Qubit mix to insure that all 2μL is dispensed. Close and label all of the sample tubes.
Vortex all tubes for 5 solid seconds.
Incubate all tubes for 3 minutes in the dark (drawer or cabinet).
Turn on the Qubit device. Select the appropriate kit. Click “yes” that you would like to use new standards. We recommend recalibrating instrument anytime you have the chance. Insert standard 1, close the lid, and press “read”. Insert standard 2, close the lid, and press “read”.
Insert your first sample and press “read”. After reading, press “calculate stock concentration” and adjust the sample volume to how much sample you have added in μL and the units to ng/μL. Record the quantity and click “save”.
Repeat step 12 for all of your samples.
Supplies