# Calibration Procedure --- tags: respirometry, macmanes title: Calibrating the FMS --- ### To work gas tanks: - Open valve, connect tube and check flow rate before connecting to the FMS - flip uper left switch up so it is at baseline, calibrate system (see below) - Close tank, depressurize regulator for storage and turn off ### Calibrate/Setting the water vapor span: - Before hooking any tubes up to the airtank, check that air and NOT water flowing through tubing - Flip the upper left switch to ‘baseline’ (instead of ‘chamber’) on FMS - Hook the a tube from FMS baseline (which is the really long tube) to the tank that contains no CO2 and let it run for 5 minutes. - You should see the water vapor values start to decrease towards zero. ### How to check the accuracy of the upper H2O span: - Use the factory set upper limit - no action required ### To zero the water vapor sensor - turn "mode dial" twice clockwise until you see "adjust zero" - then turn adjust dial (next to teh red one) to sent to zero. - hit enter (the red button below the water vapor display) ### Calibrate/Setting O2 span: - You will need a tank of 21% O2, 0% CO2, the rest Nitrogen - You only need to set the upper span of this - Tank has already been connected gas from stap 1 above. - Flow needs to be at least 250, you should be able to feel air but it should not be "blasting" - Observe the O2 reading on the FMS (it might be wrong). - to chagne the value of 02, turn the "mode dial" 3 times until it reads "fixed span O2". - Turn the adjust dial until you get to 20.9%. - Hit enter. - done ### Set CO2 zero using the current tank: - observe CO2 reading on gas analyzer (it should be zero but this is what we are adjusting) - turn "mode dial" to "zero CO2!" - hit enter - hit enter again to confirm. Display reads "hold please, this had better be CO2 free air" - done ### Set CO2 1% using the other tank - You will need a tank of 1% CO2 balanced with air - Turn on the gas very low before hooking the tubes up to baseline - Connect to tank - wait a 3 minutes for things to stabalize. - turn "mode dial" to "Adjust CO2 span" - turn adjust until you are at 1% - hit red enter button. ### change switch in upper left on FMS to computer - confirm that O2 reads about 21%