--- tags: cage testing, respirometry, macmanes --- # Cage testing Jan28 #### Empty cages ![](https://i.imgur.com/7nNCKyu.png) - Flow rate 1000 - is this measurement within the SE. E.g., is it as noisy as we expect? #### One mouse in cage ![](https://i.imgur.com/Gd8PI5M.png) - Flow rate 1000 - beautiful signal of 1 mouse #### The residual effect of 1 mouse in cage ![](https://i.imgur.com/7Qclkl7.jpg) - Flow rate 1000 #### Mouse, then invisi-mouse ![](https://i.imgur.com/NjzqZeb.jpg) - Flow rate 1500, how fast does invisimouse signal go away? - Signal does diminish, but not that fast #### One mouse Flow rate = 1500 then 1800 ![](https://i.imgur.com/ZHA4Fne.jpg) #### One mouse Flow rate = 1800 - three cages - BEAUTIFUL. Mouse is in cage 0. Baseline measurements are as expected. Returns to Baseline in cage 1 (which also does not have a mouse), then we see another consistent baseline and the same pattern for the animal on the second measurement ![](https://i.imgur.com/R9klgAT.jpg) ![](https://i.imgur.com/OJhyKz9.png)